ChloramphenicolRapid Test Strip is a one-step lateral flow immunochromatographic assay that determines a qualitative level for the presence of Chloramphenicol and is intended for use in fresh
milk, sterilized milk (pasteurised milk, UHT milk), and whole milk powder. The total test canbe finished within 15 to 20 minutes, this method is suitable for the rapid on-site test.
Limit of Detection:
0.1 μg/kg(μg/kg=ppb)
Interpretation of the Results
A color line always appears in the upper section of the test strip to indicate that the test strip is working properly. This line is the Control Line (C). A line in the lower section of the test
strip indicates the test results. This line is the Test Line (T).
Negative Sample (-):2 lines are visible. The intensity of test line (T) is greater than or equal to control line (C) indicates the sample contains Albendazole less than limitations. (Please see
image above).
Positive Sample (+): The intensity of test line (T) is less than control line (C) or only control line (C) is visible. This indicates the sample contains Albendazole greater than or equal to the
limitations.
Invalid results:
If there is no line in control zone, the test is invalid and the sample should be re-tested by using a valid test strip