Electrical parameters | |
---|---|
Accuracy | 0.05%; 0.02% |
Power Supply | One Phase AC 180-265V or 85-150V, frequency 50/60Hz. |
DC Voltage Output | |
Range |
100V, 350V, 500V, 700V (max 800V); 65V, 250V, 500V, 750V, 1000V (max 1150V) |
Adjustment range | (0-120)%RG(1) |
Adjustment fineness | 0.01%RG, 0.1%RG, 1%RG, 10%RG as optional. |
Stability | 0.01%/120s |
Distortion | 0.1% (Non-capacitive load) |
Output load | 20VA or 50VA |
Full load regulation rate | ≤±0.01%RG |
Accuracy | ±(0.03%RD+0.02%RG) |
DC Current Output | |
Range |
75mV, 2A, 20A, 50A, 100A; 0.001A, 0.005A, 0.02A, 0.05A, 0.2A, 1A, 3A, 10A, 20A, 50A, 100A, 500A (max 600A) small signal 2.5mV, 20mV, 100mV, 1V, 5V |
Adjustment range | (0-120)%RG |
Adjustment fineness | 0.01%RG, 0.1%RG, 1%RG, 10%RG as optional. |
Stability | <0.01%/120s |
Distortion | ≤0.1% (Non-capacitive load) |
Output load | 150VA or 1200VA |
Full load regulation rate | ≤±0.01%RG |
Accuracy | ±(0.03%RD+0.02%RG) |
Externally attached shunt | 100mV |
Accuracy | ±(0.06%RD+0.04%RG) |
DC Power Output | |
power output stability | <0.01%RG/120s |
power accuracy | ±(0.03%RD+0.02%RG) |
DC Power Energy Measurement Error | |
power energy | ±(0.03%RD+0.02%RG) |
Power Pulse Output | |
Power Pulse Output | 20KHz (Full-Range) |
Power Pulse Input | |
Energy pulse input | 20KHz |
Ripple output | 20-500Hz ripple output |
Communication port | USB port, RS232, 10/100M LAN |
Standard | |
Standard | JJG842-2017; JJG1085-2013; GB/T 33708-2017; IEC60664-1:2007 |
Safety | |
Isolation protection | IEC 61010-1:2001 |
Measurement Category | 1000 V CAT III |
Degree of protection | IP20 |
Declaration of conformity | CE certified |
Mechanical parameters | |
Dimensions (W×D×H) (mm) | 460x460x175; 630x600x450 |
Weight (kg) | 20; 55 |
Environmental conditions | |
Ambient temperature | -10°C to +40°C |
Relative humidity | 35%-85% |