Summary
This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93
Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.
I. Main Technical Features
1. Single chip microcomputer control technique.LCD screen displays.English man-machine interaction interface.The heating power is continuously step-less adjustable.The rate of
heating up is exact.Easy setting for parameters.
2. The LCD screen has prompting menu,prompting input operation interface.It shows the setted parameters,real-time temperatures and other test parameters.Press the “record”button
when the flash point appears.The screen will show and reserve the flash point.
3. The structure is newly design and small.Stainless steel table-board looks clean and beautiful.It can be used as long as equipped with gas or other civilian
fuels.It is easy to use and test results exact.
II. Main Technical Specifications
1. Power supply:AC( 220±10% )V, 50Hz.
2. Heating device: (1) The furnace is silicon carbide material. Power is 600W.
(2) The heating power is adjustable from 0 W to 600W.
3. Heating rate: Procedure A: (5~6)℃/min
Procedure B :(1~1.6)℃/min
Automatically controlled
4. Stirring device:(1)Stirring motor: BYGH101 stepping motor
(2) Driving
mode: flexible shaft
(3) Shaft size:
8mm×40mm
5. Stirring rate: (1) Procedure A: (90~120)RPM
(2) Procedure B: (250±10)RPM
6. Igniting device: (1) Igniting source: gas(or other civilian fuels,the same below)
(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
7. Temperature testing probe:PT100
8. Ambient temperature:≤ 35℃
9. Relative humidity:≤ 85%
10.Maximum power consumption:650W