II.Main technical specifications and parameters:
1. Titration mode: coulometric titration controlled by a microcomputer
2. Measurement range: 1 μg~100 mg (Representative value is 10 μg~10 mg)
3. Electrolysis control: It can control the electrolysis current automatically (the maximum current is 400 mA)
4. Titration speed: 2 mg per minute (maximum)
5. Accuracy:
(1)10 μg~1 mg: ±3 μg (not including sample filling error)
(2) It will be 0.3% for more than 1 mg (not including sample filling error)
6. End point: data will be shown on the instrument, buzzer will alarm, and the indication light of end point will be on
7. Date: Year Month Day Hour Minute Second
8. Printer: 16 characters pin printer, width of paper 44 mm
9. Power supply: AC 220V±10%, 50 Hz±5%
10. Power: 60 W
11. Ambient temperature: 5°C~40 °C
12. Relative humidity: ≤85%
13. Size: 290mm×380mm×120mm
14. Weight: 9 kg
Brief-Introduction The GD-2122B automatic oil moisture analyzer is designed and made as per national standard of China GB/T 7600 Standard Test Method for Water Content of Transformer Oil in Service (Coulometric Method), industrial standard of China SH/T0246 Standard Test Method ...
Place of Origin:China Model Number:KF-837 Brand Name:Kanfot Place of Origin:Guangdong China (Mainland) Certification:Ce.ROHS Type:Skin Analyzer Repeated tolerance: 1.5%, Graduation values: 0.1%. Moisture range:10~65%l Oil range:8~32%l ...