II.Main technical features
1.Designed as per GB/T 5208-2008 Determination of flash point - Rapid equilibrium closed cup
method. It can realize purpose of testing closed cup flash point rapidly and at low temperature.
2.The instrument adopts semiconductor refrigeration device and external water cooling device.The size is small and cooling rate is rapid.Easy to operate.
3.Test sample needed is less.2ml for each time(4ml for solid sample and semi-solid sample)
4.The determination procedure is automatically completed except ignition.Test results can be printed automatically.
III.Main technical specifications and parameters
1. Power supply: AC(220±10%)V, 50Hz
2. Flash point detecting range: (50~100)°C
3. Precision of determination:
Absolute difference value between two determinations is lower than 2°C(same operator)
Absolute difference value between two determinations is lower than 3°C(different
operator)
4. Accuracy of temperature control: ±0.5°C
5. Igniting device:Electric igniting gun
6. Cooling mode:Semiconductor(with cold water cycle at external connection)
7. Overall Power Consumption:Not higher than 300W
8. Ambient temperature:(5~30)°C
9. Relative humidity:(30~80)%
10. Overall dimension:370mm×280mm×280mm(L×W×H)
Notification: Our company can customize instruments with different ranges of low-temperature flash point according to customer's need.
Purpose This GD-3536A Automatic Cleveland Open Cup Flash Point testing equipment is designed and made as per GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points- Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by ...
Purpose The Open Cup Flash Point testing equipment is designed and made as per the National Standard GB/T3536 “Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Methods), and T0651 “Asphalt Flash Point and Fire Point Test (Cleveland Open ...