Types of Medical Alarm Systems
*Also available in Retro-Fit models.
*Master alarms monitor source equipment, such as: Air Compressors, Vacuum Pumps, High/Low Pressure Switches, Manifolds and other pipeline devices. Up to 60 devices can be monitored on each alarm.
Pressure: HIGH - 60psi / LOW - 40psi / Normal - 50psi
Nitrogen: HIGH - 195psi / LOW - 140psi / Normal 165psi
Vacuum: LOW - 12 inHg / Normal - 20 inHg.
A trend bar is provided to show the direction of the gas/vacuum pressure.
The ALERT-2 medical gas alarm is used by hospital personnel to monitor gas pressures and vacuum levels of the medical gas pipeline. All gases or vacuum levels are displayed with large red LED's for clear visibility, readable even in dim lighting conditions.
Pressure: HIGH - 60psi / LOW - 40psi / Normal - 50psi
Nitrogen: HIGH - 195psi / LOW - 140psi / Normal 165psi
Pressure: HIGH - 60psi / LOW - 40psi / Normal - 50psi Nitrogen: HIGH - 195psi / LOW - 140psi / Normal 165psi |
Magnetic gas medical waste pyrolysis device. Used for medical waste, garbage collection, incinerator, garbage disposal. Wood packing. Installation length: 6.8 meters. Equipment width: 3 meters. Equipment height: 4.5 meters. Building width: 7 meters. Building length: 12-15 ...
Come From Hangzhou fireprint Technology Co., Ltd.
Oxygen Machine Uses We have the most authoritative and professional team to provide you with solutions such as technical design, pre-sales service, and after-sales support. 01 Medical and Hospital Oxygen concentrators are widely used in treating cardiovascular diseases, ...
Come From SHENYANG CANTA MEDICAL TECH. CO., LTD
The effect after using your hydrogen inhalation 1 week: 1. Improve energy, fight fatigue. 2. Improve the quality of sleep. Easier to fall asleep and sleep deeper. 3. Make the gastrointestinal tract healthier and improve constipation. The effect on the female after using it 2 ...
Come From Hydrogen Special Inc.,