5. DSH128 Gas Chromatograph 气相色谱仪 (无图片)
Gas Chromatograph , Gas Analyzer , Chromatograph
I.Technical characteristics:
1. Gas path control technology using ECD. Flow pressure within a set range, arbitrarily set, to automatically adjust a throttle function.
2. High precision program temperature control system , the oven temperature could be controlled at ± 0.02 ℃.
3. With complete system self-test function and automatic fault recognition.
4. Have a system of automatic parameter memory function, all parameters can be automatically controlled by workstation, ready to call the store analytical conditions.
5. TCD filament gas cut off with automatic protection function
6. With twenty external event control functions.
7. The network data communication and remote control.
8. Valve configurations with a variety of solutions.
9. High reliability and high sensitivity detector. Column oven:
10. Temperature range: room temperature + 5 ℃ ~ 400 ℃
11. Control accuracy: ± 0.02 ℃
12. Temperature programming: 9 bands / 10 platform
13. Maximum heating rate: 60 ℃ / min
14. Single-channel compensation algorithm
15. Automatic cooling injector:
16. You can load three injector
17. Injector unit: split / splitless injector detector:
18. You can load three detectors, gas detectors using PPC control
19. Hydrogen flame ionisation detector: Temperature range: + 7 ℃ ~ 400 ℃ detection limit: ≤ 3 × 10-12g / s sample: C16
20. Micro thermal conductivity detector (μ-TCD): Temperature range: + 7 ℃ ~ 400 ℃ Sensitivity: ≥ 10000mV × .ml / mg sample: C16
21. Electron capture detector (ECD): Temperature range: 7 ℃ ~ 350 ℃ sources: 63Ni10 ~ 12mci minimum detectable amount: 8 × 10-14g/ml sample: г666
22. Flame photometric detector (FPD): Temperature range: + 7 ℃ ~ 350 ℃ detection limit: ≤ 3 × 10-12g / s (P) ≤ 3 × 10-11g / s (S) sample: formaldehyde parathion carrier gas flow rate:
23. Carrier gas flow control (EFC) and the head pressure control (EPC)
24. Pressure setting range: 0 ~ 100Kpa
25. Total flow setting range: 0 ~ 200ml/min (N2)
26. Auxiliary gas: 0 ~ 100ml/min (H2); 0 ~ 500ml/min (Air)
I.Product Introduction The chromatograph system newly designed with high-performance HPLC. Instrument (base model) is formed by the high-pressure infusion pump, injector separation system, UV / VIS detector . II.Features: 1. Adopts microcomputer dual-head pump complementary ...
I.Features and performance: 1. The newly designed GC126 microcomputer temperature control system, using the international advanced control technology. High temperature accuracy (better than ± 0.1 ℃) ,reliability . 2. Keyboard settings various control and parameters, the ...
3. DSH122 GC112 Gas Chromatograph 气相色谱仪 G3 Gas Chromatograph , Gas Analyzer , Chromatograph GC122专用选配检测器:见网站 I.Key Features: 1. The newly designed high-performance HRGC. 2. The instrument equipped with a dual FID detector, optional GC122-TCD ...
I.Characteristics 1.Brand-new designed microcomputer temperature control system of world advanced control technology. High accuracy of temperature control ( optimum:0.05C), high reliability and anti-interference. 2.Control and operating parameters(including for detector) set up ...
I.Features: 1. Microcomputer temperature control with high precision (better than ± 0.1 ℃). 1. All control and application parameters, including operation parameters of the detector, can be set up through keyboard. Functions, such as self-diagnosis, switch-off protection, ...
Usage:Power Model Number:Mykos Brand Name:Morgan Place of Origin:Canada a portable Micro Gas Chromatograph that can measure separately the seven fault gases found in transformer oil when a problem is ...