I.Features
The product is completely controlled by the PC , can flexibly matching flame and graphite furnace atomizer . The unique optical and mechanical design, safe and convenient flame systems, advanced graphite furnace temperature control technology, optional background deduction techniques, and the convenient function offered by the workstation to suit your exact measurement results for automation . With up to 500 or more data storage, and power off storage capabilities.
II. Technical Parameters
1.Wavelength range: 190 - 900nm
2.Spectral bandwidth: 0.1, 0.2, 0.4, 1.0 and 2.0nm
3.Accuracy of the wavelength: ±0.15nm
4.Receptivity of the wavelength : ≤0.04nm
5.Base line stability:≤0.002A/30min (Cu)
6.Characteristic viscosity:0.02μg/ml/1% (Cu)
7.Checking limit:0.004μg/ml (Cu)
8.Preciseness :0.5%
9.Grating:1800 lines/mm
10.Inflamer:All-metal titanium burner
11.Atomizer :Effective glass atomizer
12.Lamp stand:8
13.D2 background correction ability :When the background is 1 A, the background ability should be deducted not less than 50 times; self absorption background deduction method
14.Size and weight :700mmx550mmx450mm 75Kg
15.Power source:220V±22V AC 7000W
(I) Flame system
Acetylene air burner :100mm
Nitrous oxide acetylene flame head:50mm
Ignition dynamic baseline drift:≤0.006A/30min
(Cu) Characteristic viscosity :≤0.025μg/ml/1%
Related standard deviation of the accuracy:≤0.5%(Cu, absorbance>0.8A) (detection limit Cu≤0.008μg/ml)
Safety system:Can automatically cut off the gas when the pressure is not enough, the power is off, flameout and unconformity of the burner.
I.Features 1.This is a kind of Zeeman background correction AAS which used Zeeman detection system designed by ourselves (patent No.: 27523). 2.It has higher detection efficiency, better signal and noisy ratio, therefore, the instrument with Zeeman system is better than same ...
I.Technical Features 1.Deuterium lamp background correction system. 2.Selection in double lamp and lamps auto conversion device. 3.Auto-wavelength scanning setting, auto-lamp current, auto-voltage and auto-regulation of energy. 4.Flame monitoring protection system. 5.Titanium ...
I.Features 1.Deuterium lamp background correction system. 2.Grating 1800L/mm, short focal length 270mm, better signal and noisy ratio. 3.Auto-wavelength scanning,auto-lamp current,negative auto-voltage and auto- energy balance of cathode lamp and deuterium lamp. 4.High ...