(1)Low Stray Light
The stray light of V-5800(PC) is below 0.05%T.The monochomator is totally sealed and the optical surfaces are protected with a silicon dioxide membrane.The excellent optical
system and electric circuit design ensures the very low stray light.
(2)Large LCD display
Models V-5800(PC) adopt 128*64 Dots LCD as its display system. Thus makes the results easy to read. And the Standard Curve can be displayed on it clearly,Also,it can display 5
groups of data per screen.
(3)Large Data memory space
It can store 200 groups of data and 200 standard curves at the same time.Press arrow key to scroll,easy to reload.
(4) Can upgrade to scanning modle
The instrument can be controlled by Metash’s scanning software UV-Professional developed based on Windows® through the USB port.The software expands the function to
Quantitative,Kinetics,Multiwavelength and Wavelength Scanning.
Specifications
Model DSH-UV-5800(PC)
Wavelength Range 320-1100nm
Bandwidth 2nm
Wavelength Accuracy ±0.5nm
Wavelength Repeatability 0.2nm
Wavelength Setting Auto
Photometric Accuracy ±0.3%T
Photometric Repeatability 0.2%T
Photometric Range -0.3-3A,0-200%T,0-9999C
Stability ± 0.002A/h @ 500nm
Baseline Flatness ± 0.0015A/h
Stray Light ≤0.05%T
Data Output Port USB
Printer Port Parallel Port
Display 128*64 Dots LCD
Lamps W Lamp
Detector Silicon Photodiode
Power Requirements AC 220V/50Hz or 110V/60Hz
Dimension 460*360*225mm
Weight 18kg