GD-265H Kinematic Viscosity Test Machine
Brief Introduction
Main technical specifications and parameters
1.Power supply: AC 220 V±10%, 50 Hz
2.Heating power of water bath: 1700 W
3.Temperature range of water bath: room temperature~100.0 °C
4.Temperature controlling accuracy: ±0.01 °C
5.Cubage of water bath: 20 L
6.iming range: 0.0 s~9999.9 s
7.Timing accuracy: ≤ ±0.05% in 60 minutes
8.Sample quantity: four pieces of capillaries are installed at the same time
9.Stirring motor:
(1) Power: 6 W (2) Rotation speed: 1200 RPM
10.Operation environment:
(1) Ambient temperature: -10 °C~+35 °C (2) Relative humidity: ≤85%
11.Temperature sensor: industry platinum resistance, scale number is Pt100
12.Total power consumption: ≤ 1800 W
13.Capillary viscometer: 7 pieces in total. Inner diameters are 0.6mm, 0.8mm, 1.0mm, 1.2mm, 1.5mm, 2.0mm, and 2.5mm respectively.
Summary This instrument is designed and made as per the industry standard of People’s Republic of ChinaSY/T5651 Technical Condition of Petroleum Products Kinematic Viscosity Tester, GB/T 265 Petroleum products-Determination of kinematic viscosity and calculation of dynamic ...
Summary The instrument is designed and made as per the National Standard of People’s Republic of China GB/T 265 Petroleum products-Determination of kinematic viscosity and calculation of dynamic viscosity andGB/T1632-93 Test Methods for Viscosity Number and Limiting Viscosity ...
I.Brief Introduction The instrument is suitable to determine the kinematic viscosity of liquid petroleum products (Newton liquid) under a fixed temperature environment as per GB/T 265 Standard Test Method for Kinematic Viscosity and Calculation Method for Dynamic Viscosity of ...